Intraos 70 ошибка a02

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    1/52

    IntraOs 70X-ray EquipmentService & Installation Manual

    Blue X Imaging Srl Via Idiomi 1/8-33 20090 Assago ITALY e-mail
    [email protected]

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    2/52

    IntraOs 70 Service & In

    2/52 69 500 00210

    1 Medical device for emission of ionizing radiation on re2
    System turned ON and ready

    3 Irradiation

    4 Alarm

    5 Exposure time in s6 Manual decrease of exposure time

    7 Manual increase of exposure time

    8 Patient size large

    9 Patient size small

    10 Maxillary incisor11 Maxillary canine or premolar

    12 Maxillary molar

    13 Mandibular incisor

    14 Mandibular canine or premolar

    15 Mandibular molar

    16 Bite-wing premolar

    17 Use of digital detector

    18 Radiation exposure pushbutton

    Control panel and exposure pushb

    1 2 3

    56

    8

    9

    10 11 12

    13 14 15

    18

    1

    stallation Manual

    Version 4.0

    uest

    utton

    4

    7

    16

    17

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    3/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    SET-UPMICROSWITCHES

    KEYBOARDCONNECTOR

    BACK UP TIMTEST POINT &

    MICROPROCESSORWATCH DOG LED

    FUSE 4T 200 mA

    5 x 20

    PS24

    FUST 200

    5 x

    OUTPUTLINE TO

    TUBEHEAD

    INPUTMAINS LINENEUTRAL

    & LIVE

    JUMPERFUSE 2

    MAINSSWITCH

    FUSE 1

    & FUSE 2T8A 6.3×3

    Layout Power Board Timer Aut

    3/52

    POWERSUPPLY5 V LED

    RED

    WERPPLYV LED

    E 3mA

    20

    HANDSWITCHSOCKETS

    HANDSWITCHJUMPERS

    Set

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    4/52

    IntraOs 70 Service & In

    4/52 69 500 00210

    Blue X Imaging S.r.l.

    Via Mario Idiomi 1/8-3320090 Assago ITALY

    tel.+39.0245712171fax +39.0245703385

    e-mail [email protected]

    IntraOs 70 Dental X-ray EquipmentService & Installation
    Manual English Editi

    Version 4.0 March 2008

    Printed 30/05/2008 12:11:00Code 69 500 00210

    Blue X Imaging is committed to Total Qualit

    Thank you for notifying us any error found in the do

    stallation Manual

    Version 4.0

    n

    .

    cument.

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    5/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Table of Contents

    1. INTRODUCTION
    ……………………………………………….1.1 Purpose
    ……………………………………………………….1.2

    Equipment Classification ………………………………1.3
    Applicable Standards ……………………………………1.4
    Environmental Data ……………………………………..1.5
    Obligations of the Installer …………………………….1.6
    Warning………………………………………………………1.7
    Demonstration
    ……………………………………………..

    2. TECHNICAL
    DATA……………………………………………2.1 System
    Supply …………………………………………….2.2 Tube
    Housing Assembly ……………………………….

    2.3

    Beam Limiting Device ………………………………….2.4
    AutoSet Timer
    ……………………………………………..2.5 Mechanical
    Suspension System ……………………..2.6 Weights
    ………………………………………………………

    3. ASSEMBLY AND INSTALLATION ……………………3.1 Wall
    Mounted Systems …………………………………3.2 Mobile
    Systems ……………………………………………

    4. MAINTENANCE
    ………………………………………………..4.1
    Maintenance of the Wall Support …………………..

    4.2

    Maintenance of the Extension Arm ………………..4.3
    Maintenance of the Folding Arm ……………………4.4
    Maintenance of the Tube-Head ………………………4.5
    Maintenance of the Beam Limiting Device ……..4.6 Maintenance of
    the Timer ……………………………..4.7 Checking Filament
    Pre-Heating Time …………….4.8 Special Service Functions
    ……………………………..

    5. MEASUREMENTS
    ……………………………………………..5.1 Line
    Voltage
    ………………………………………………..

    5.2

    Anode Voltage KVp…………………………………..5.3
    Anode Current mA …………………………………….5.4
    Exposure Time
    …………………………………………….5.5 Leakage
    Radiation ……………………………………….5.6 Earth
    Resistance …………………………………………..5.7
    Earth Leakage
    ……………………………………………..

    6. Spare Parts
    ………………………………………………………….6.1
    Tube-Head
    …………………………………………………..6.2
    Folding Arm
    ………………………………………………..

    6.3

    Extension Arm
    …………………………………………….6.4

    Wall Support
    ……………………………………………….6.5 Mobile
    Base ………………………………………………..6.6
    AutoSet Timer
    ……………………………………………..

    5/52

    ……………………… 7

    …………………….. 7

    …………………….. 7

    …………………….. 7

    …………………….. 7

    …………………….. 8

    …………………….. 8

    …………………….. 8

    ……………………… 9

    …………………….. 9

    …………………….. 9

    …………………… 10

    …………………… 10

    …………………… 11

    …………………… 11

    ……………………. 12

    …………………… 12

    …………………… 25

    ……………………. 28

    …………………… 29

    …………………… 29…………………… 30

    …………………… 34

    …………………… 34

    …………………… 34

    …………………… 35

    …………………… 35

    ……………………. 36

    …………………… 36

    …………………… 36…………………… 36

    …………………… 37

    …………………… 38

    …………………… 38

    …………………… 39

    ……………………. 40

    …………………… 40

    …………………… 41

    …………………… 42

    …………………… 43

    …………………… 44

    …………………… 45

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    6/52

    IntraOs 70 Service & In

    6/52 69 500 00210

    Appendix A System Components
    …………………………………….Appendix B Icons
    ………………………………………………………….Appendix
    C Exposure Table
    ……………………………………………Appendix D Alarm
    Conditions ………………………………………..Appendix
    E Identification Labels
    …………………………………….Appendix F Cooling
    Curves ……………………………………………

    Electrical requirements: see section 3.1.3 at page 13.

    To connect the timer see section 3.1.10 at page 19.

    INPUTPOWERLINE

    SWITCH

    ELECTRONICTIMER

    WALLSUPPORT

    X-RAYHEAD

    IN OUT

    IN

    GROUND

    HAC

    HAND-SWITCHSOCKET TST1

    FUSE F1

    FUSE F2

    OUT

    ZERO OHMRESISTOR

    FUSE

    HAND-SWITSOCKET TS

    NEUTRAL

    LIVE

    stallation Manual

    Version 4.0

    …………………. 46

    …………………. 47

    …………………. 48

    …………………. 49

    …………………. 50

    …………………. 51

    D-SWITCHNNECTIONUMPERS

    4

    H2

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    7/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    1. INTRODUCTION

    1.1 PurposeThe IntraOs 70 X-ray Equipment is design to fulfil
    the nee

    radiography in the general dental practice.The system can be
    configured for wall, unit or mobile soluttypes of timers and
    tube-heads.

    The features of the system make it easy to use, and grant lowith
    minimum maintenance.

    1.2 Equipment Classification IEC: IntraOs 70 is a Class I, type
    B equipment

    FDA: IntraOs 70 is a Class II equipment (21 CFR 872-

    1.3 Applicable StandardsThe IntraOs 70 system configurations,
    all equipped with thcomply with the following standards.

    IEC 601-1 General requirements for safetyIEC 601-1-2
    Electromagnetic compatibilityIEC 601-1-3 General requirements for
    radiation pr

    diagnostic X-ray equipment

    IEC 601-2-7 Particular requirements for the safetygenerators of
    diagnostic X-ray genera

    IEC 601-2-28 Particular requirements for the safetyassemblies
    and X-ray tube assembliediagnosis

    21 CFR 1020.30 Diagnostic x-ray systems and their m21 CFR
    1020.31 Radiographic equipment

    1.4 Environmental Data

    Applicable ranges of temperature, humidity, and atmosphereported
    here below both for operation and transport condit

    Ambient TranTemperature from 10 to 40 C froRelative Humidity
    from 30 to 75%Pressure from700 to 1060 hPa from

    7/52

    s for intra-oral

    ions and different

    g useful life

    1800).

    AutoSet timer,

    tection in

    f high voltageorsf X-ray sourcefor medical

    jor components

    ic pressure areions.

    sport & Storage20 to +50 C10 to 90%500 to 1060 hPa

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    8/52

    IntraOs 70 Service & In

    8/52 69 500 00210

    1.5 Obligations of the InstallerObligations of the Installer
    are:

    A To make sure that the line voltage specified by the

    Manufacturer of the equipment is available and within
    thspecified range.B For safety reasons verify that a proper switch
    is availabl

    from line voltage supply when needed during installatioC Install
    and test the equipment with due diligence accordi

    installation instructions from the Manufacturer.D To provide the
    Operators Manual to the User.

    1.6 Warning

    X-ray equipment produce ionising radiation that may be
    harproperly controlled. It is therefore recommended that the
    equoperated by trained personnel only in accordance with existi

    Even if compliant to specifications of electromagnetic
    comprecommended not to use the equipment in presence of
    externelectromagnetic fields, such as those generated by cellular
    pmight interfere with the electronic circuits of the system.

    1.7 DemonstrationIn order to use of the system for demonstration
    purposes radihas to be inhibited by disconnecting the supply cables
    to thethe wall support or into the timer.

    Cables to be disconnected are those leaving the connection
    btube-head and not those coming in from the timer.

    Make sure that the disconnected cables are properly
    insulateundesired contacts with live points.

    This task has to be done by trained personnel only to avoid
    telectrical shock.

    DISCONNECTHERE

    INPUTPOWERLINE

    SWITCH

    ELECTRONICTIMER

    WALLSUPPORT

    X-RAYHEAD

    IN OUT

    stallation Manual

    Version 4.0

    e

    to disconnect.g to the

    ful if notipment beg law.

    tibility, it isalones, which

    ation emissionube-head into

    ock towards the

    to prevent

    e risk of

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    9/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    2. TECHNICAL DATA

    2.1 System SupplyLine Voltage 115 V (from 99 V to 137 V in

    depending on tube housing a230 V (from 198 V to 275 V idepending
    on tube housing a

    Line VoltageWorking Range

    Limited to the working rangefrom108 V to 132 V for typefrom 207
    V to 253 V for type

    MaximumLine Current

    8 A at 115 V5 A at 230 V

    Fuse 8A slow blow at 115 V,

    5 A slow blow at 230 Vfor two-phase supply or on mthe second
    fuse to be activat

    Line Frequency 50/60 Hz 2 HzLine resistance 0.4 Ohm at 115
    V,

    0.8 Ohm at 230 V

    2.2 Tube Housing AssemblyNominalLine Voltage

    120 V for type 93 200 01300,230 V for type 93 200 01700

    Line VoltageWorking Range

    120 V 10% for type 93 200230 V 10% for type 93 200

    Anode voltage70 kVp 8% at nominal line66 kVp 8% at nominal
    line74 kVp 8% at nominal line

    Anode current7.0 mA 15% at at nominal li5.3 mA 15% at nominal
    line8.3 mA 15% at nominal line

    Maximum load 70 kVp, 7 mA, 3.2 sX-ray insert 3 electrodes, grid
    control acti

    models: OCX/70-G, RFG070Focal Spot 0.8 (EN
    60336:1995-04)Inherent filtration > 2.5 mm AlDuty Cycle
    1/30Radiation Leakage < 0.1 mGy/h a 1 m (< 11.5 m

    9/52

    sub-ranges

    ssembly)sub-ranges

    ssembly)f the THA:

    3 200 01300,3 200 01700

    obile system,d by jumper cut

    01300,01700oltageoltage 10%oltage+ 10%ne voltagevoltage
    10%voltage + 10%

    n

    /h a 1 m)

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    10/52

    IntraOs 70 Service & In

    10/52 69 500 00210

    2.3 Beam Limiting Device

    Round BLDMetal cone with near-focus seFocus skin distance (FSD)
    8.2Circular radiation field

    size 2.35 diameter (6 cm)

    Rectangular BLDMetal body with near-focus seFocus skin distance
    (FSD) 8.2Rectangular radiation fieldsize 1.38×1.81 (3.5×4.6 cm)

    2.4 AutoSet TimerLine VoltageWorking Range

    115 V 15% for type 93 300230 V 15% for type 93 300

    Exposure factor Exposure time in s,18 steps from 0.06 s to 3.2
    s0.06 0.060.08 0.080.10 0.10Time in seconds is converted

    mains pulses with 1 pulse preat 50 Hz, 16.6 ms at 60
    Hz.Precision 0.02 s or 5% (whichever thExposure factor setting
    Automatic setting through too

    selection and patient size, fortraditional film or digital
    sensosetting moving up or down thplus or minus keys.

    Irradiation signal Yellow light on hand-switch a

    panel plus acoustic buzzerPreheating time Set at installation
    (typical valuHand-switch Hand-switch with 3 m coiled c

    remote mounting possibilityOverall size 6 (15 cm) width, 9 (24
    cm)

    cm) depthOther features Microprocessor controlled fun

    Film speed setting3 s minimum waiting time

    Energy management for coolOptional kit with 33 ft (10 m)remote
    hand-switch mounting

    stallation Manual

    Version 4.0

    tion(21 cm)

    tion(21 cm)

    02000100

    R10 scale)0.060.080.10

    to number of

    cision (20 ms

    greater)h typeuse withr, or manualscale with

    d on control

    e 180 ms)ord,

    height, 3 (9

    ctionality

    down timeable for

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    11/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    2.5 Mechanical Suspension SysteWall Support 4.72 (12 cm)
    width,

    9.45 (24cm) height,3.54 (9 cm depth)

    Extension arm, length Short: 17 (45 cm),medium:27 (70 cm),long
    35 (90 cm),extra long 43 (110 cm)

    Folding arm;useful reach

    55 (140 cm) with short exten65 (165 cm) with medium ex72 (185
    cm) with long exte80 (205 cm) with extra long

    Mobile Stand 29 (74 cm) width,

    42 (107 cm) height,24 (62 cm) depth65 (165 cm) total heath
    with

    2.6 WeightsTimer 3.7 lb (1.7 kg)Tube-head: 14.1 lb (6.4 kg)Round
    BLD 0.22 lb (0.1 kg)

    Rectangular BLD 0.44 lb (0.2 kg)Folding Arm: 15.4 lb (7 kg)Short
    Ext. Arm 7.7 lb (3.5 kg)Medium Ext. Arm 9.3 lb (4.2 kg)Long Ext.
    Arm 11 lb (5 kg)Extra long Ext. Arm 12.7 lb (5.8 kg)Wall Support:
    2.9 lb (1.3 kg)Mobile Stand 55.1 lb (25 kg)

    11/52

    sion armension arm

    nsion armxt. arm

    folding arm

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    12/52

    IntraOs 70 Service & In

    12/52 69 500 00210

    T1

    B1

    3. ASSEMBLY AND INSTALLATION

    3.1 Wall Mounted Systems

    3.1.1 Unpacking

    Unpack the components of the system and checkthe following:

    A Each item is in good conditions and was notdamaged during
    transportation.

    B All the items for the desired systemconfiguration are
    available.

    C The line voltage on the labels of timer andtube-head are
    corresponding to the existing

    local line voltage.3.1.2 Structural Requirements

    The wall support has to be mounted in a convenient positionside
    of the chair or on back wall (head of the patient).

    The maximum useful reach is of 80 (205 cm) from the
    wallcombining a 43 (110 cm) extension arm with a folding ar

    The wall support can be mounted with 2, 4, or 6 bolts,
    depenquality.

    A Two bolts only (top and bottom centralholes T2, B2) are used
    when there is asolid slim column (e.g. iron mounting) withweak
    sides (e.g. wooden wall). Consideringthe requested safety factor,
    each one of thetwo bolts has to withstand a load of 5470
    N,comprehensive of a safety factor, i.e. about1240 lbs or 560 kg.
    Proper screw to beselected for a solid connection to the wall.

    Classes ISO 8.8 (M 8, M 8×1, M 8×1.25) orSAE Grade 5 (5/16
    18UNC, 5/16 24UNF) are recommended.

    B Four bolts, two on top sides (T1, T3) andtwo on bottom sides
    (B1, B3) is the regularmounting for solid (concrete) wall, but
    alsoon large metal plate. Considering therequested safety factor,
    each bolt has towithstand a load of 2735 N, comprehensive

    of a safety factor, i.e. about 620 lbs or 280

    stallation Manual

    Version 4.0

    T2 T3

    B2 B3

    on left or right

    when.

    ing on wall

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    13/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    kg. Proper expansion screw to be selected for a solid cconcrete
    wall; the permissible load of each screw has t308 lbs (about 140
    kg). On solid concrete use heavy duty metal anchors.

    On hollow bricks use injection chemical fixing.C Six bolts,
    three on top (T1, T2, T3) and three at bottom

    required when the wall is not solid enough and the
    loaddistributed on more points. Considering the requested sbolt has
    to withstand a load of 911 N, comprehensive oi.e. about 210 lbs or
    95 kg.

    In case the wall is not in condition to withstand the
    indicateactions can be evaluated by adoption of reinforcing
    plates:

    D Large plate to fit vertical supports at 16 distance, withand
    one cable opening for the wall mount in the middle

    E In case of a thin (wooden) wall not solid enough, the uplate
    (counter plate) made of iron (2 mm thick) can beMake sure that the
    wall is solid enough to carry the loa

    F The use of two reinforcing iron plates of about 4 timeswall
    support, one by each side of the wall, can help whplate looks not
    adequate; additional bolts have to be ustogether the two
    plates.Make sure that the wall is solid enough to carry the loa

    Recommended bolts are listed belowRecommended Bolts

    Diameter Class CorM 8X1.25 ISO 8.8

    M 8X1 ISO 8.85/16 18 UNC SAE- Grade 55/16 24 UNF SAE- Grade
    5

    INSUFFICIENT WALL OR HARDWARE STRENGHMAY CAUSE THE WALL MOUNT TO
    PULL OUT FTHE WALL AND THE FULL SYSTEM TO FALL OTHE PATIENT OR THE
    OPERATOR CAUSINGINJURIES.

    3.1.3 Electrical Requirements

    The power line cable must be connected to the input termintimer
    to supply the timer itself and to make available powehead at the
    output terminals (OUT), upon request by the ophand switch).

    Power to the X-ray head is thus controlled by the timer (S

    13/52

    nnection to thebe greater than

    (B1, B2, B3) arehas to be

    afety factor, eachf a safety factor,

    d load, corrective

    4 mounting holes

    .e of a reinforcinghe solution..he surface of there one singled
    to keep

    .

    Section mm36.639.233.8

    37.41

    TROMTO

    als (IN) of thefor the X-ray

    erator (via the

    ITCH).

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    14/52

    IntraOs 70 Service & In

    14/52 69 500 00210

    The X-ray head is connected to the output terminals of the
    tithrough the connecting block in the wall support.

    Cabling (2 poles plus ground) to connect the power line to
    thtimer to the wall support is not provided and has to be
    procurlarge section conductors keeps to minimum the wire
    resistanrelevant voltage drop.

    Wire Type Square Section mm ResisAWG 16 1.35AWG 14 2.11

    The hand-switch too can be mounted far from the timer, prois in
    view and the sound of the buzzer can be heard.

    3.1.4 Mounting and Connecting Sequence

    A Mount the Wall SupportB Mount the TimerC Mount the Extension
    ArmD Mount the Folding/Simple ArmE Connect the Wall SupportF
    Connect the TimerG Connect the hand-SwitchH Optional Remote
    hand-SwitchI Mount and connect the Tube-headJ Mount the Beam
    Limiting DeviceK Final Tuning and Set-Up

    INPUTPOWERLINE

    SWITCH

    ELECTRONICTIMER

    WALLSUPPORT

    X-RAHEA

    IN OUT

    stallation Manual

    Version 4.0

    er (OUT)

    e timer and theed. Cable withe and the

    ance Ohm/m2x0.015×0.0094

    ided the patient

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    15/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    T1

    B1

    3.1.5 Mounting the Wall Support

    A Remove the plastic coverunscrewing the screws under thelogo
    label.

    B Use the Wall Support plate or atemplate to mark the holes on
    thewall. Please note that the timercan be mounted close to the
    WallSupport on the right side or in aremote position. Cabling
    shouldhave been laid out accordingly.

    C Make the holes in the wall according tothe applicable type of
    mounting:a Two bolts,b Four bolts,c Six bolts,d With or without
    reinforcing plate.

    D Slide the logo-strip out from the plasticcover to access the
    screws and remove itfrom the metal frame.

    E Mount the metal frame on the wall usingproper heavy-duty metal
    anchors andmake sure that the power cable entersfrom behind.

    F Secure it to the wall ensuring it is level.Improper levelling
    might cause the armto move and swing out ofposition.

    G Perform wire connection andfinal set-up complying
    withrecommended sequence of actionsreported in the following.

    15/52

    T2 T3

    B2 B3

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    16/52

    IntraOs 70 Service & In

    16/52 69 500 00210

    3.1.6 Mounting the Timer

    A Use the mounting plate or thetemplate to mark the holes onthe
    wall.

    B Slide the logo-strip out fromthe plastic cover to access
    thescrews and remove the plasticcover from the plastic frame.

    C Drill the four holes in the walland secure the timer
    makingsure it is level. Cabling shouldhave been laid out
    inaccordance to mounting

    requirements.D Perform wire connection and

    final set-up complying withrecommended sequence ofactions
    reported in thefollowing.

    3.1.7 Mounting the Extension Arm

    A Unpack the extension arm and check for completeness oB Cut,
    but do not remove, the string to be used later on to

    from the Folding/Simple Arm through the Extension ArSupport.

    C Lightly grease the support tube at base of the extension atop
    of wall mount.

    D Insert, but without tightening, the locking and the
    frictio

    stallation Manual

    Version 4.0

    parts.ull the cable

    to the Wall

    rm and insert at

    screws.

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    17/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    SAFETYSTRIP

    3.1.8 Mounting the Folding Arm

    WARNING.THE SPRINGS IN THE FOLDING ARM

    MAY CAUSE INJURY TO THEINSTALLER AS WELL AS DAMAGETO THE ARM
    ITSELF IF NOTHANDLED PROPERLY.DO NOT REMOVE BINDING UNTILWHEN
    NECESSARY AS INDICATED INTHE INSTRUCTIONS BELOW.

    A Slide the blocking ring onto the support

    tube at base of the arm.B Light grease the mounting pin at base
    of

    Folding Arm.C Secure the driving string to the end of the

    electrical cable of the Folding Arm.D Pull the other end of the
    string to drive

    the electrical cable of the Folding Armthrough Extension Arm out
    into WallMount.

    E Hold Folding Arm sections together andcarefully remove binding
    allowing thearm sections to open slowly, away frompeople.

    F Perform wire connection (into the wallSupport) and final
    set-up complying withrecommended sequence of actionsreported in the
    following.

    WARNING.NEVERCONNECTTHE CABLEOF THEFOLDINGARMDIRECTLYTO

    MAINSLINE.

    17/52

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    18/52

    IntraOs 70 Service & In

    18/52 69 500 00210

    TOFOLDINGARM

    EXTENSIONARM

    FROMTIMER

    WALL SUPPORT CONNECTIONS

    GROUNDCONNECTIONSON WALLPLATE

    3.1.9 Connecting the Wall Support

    A Connect the three wires from the Timer (additional cableblock
    and to the Folding/Simple Arm (Tube-head) as repfigure.

    B Connect the ground cable on the Wall Support to the blothe
    extension arm

    stallation Manual

    Version 4.0

    ) to the terminalorted in the next

    king bolt on

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    19/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    3.1.10 Connecting the Timer

    A Turn-off the line voltage supply line.

    B Connect the three wires from the line voltage supply
    toterminal block (IN on the left) but do not connect anyoutgoing
    wire (OUT on the right) to Wall Support.

    C Make sure that the live connector is the hot one:

    a Connect an AC voltmeter or a test light between thlive in and
    ground.b Turn line voltage supply ON. If full line voltage

    lamp lights) the wiring is correct. If not turn line vOFF,
    reverse live in and neutral in wires anmeasurement; eventually the
    full line voltage shoulbetween line and ground (test lamp
    lights).

    c Make sure no voltage is read between neutral innot check line
    voltage distribution.

    D Test the timer for full functionality keeping the
    outputdisconnected so that power is not provided to the load.

    IN

    GROUND

    HAND-SWITCSOCKET TST

    FUSE F1

    FUSE F2

    OUT

    ZERO OHMRESITOR

    FUS

    HAND-SWSOCKET

    NEUTRAL

    LIVE

    19/52

    the

    block terminal

    s measured (testltage supplyrepeat the

    d be read

    nd ground; if

    ables

    AND-SWITCHCONNECTION

    JUMPERS

    E F4

    TCHST2

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    20/52

    IntraOs 70 Service & In

    20/52 69 500 00210

    E Connect the load. Turn OFF the line

    voltage supply line,

    Connect the three

    wires of additionalcable outgoing to theFolding Arm
    (WallSupport).Live and neutral wiresthe folding arm (tube-head) can
    be interchanged.

    Table of Fuses for Timer AutoSet

    Fuse 115 V 2F1 T8 A 6.3×32 T5F2 T8 A 6.3×32 T5

    F3, F4 T200 mA 5×20 T200

    3.1.11 Connect the Hand-Switch

    The hand-switch is provided with a 3 m coiled cord the
    termiplugged in the lower left corner of the board, either
    externall

    TST1) or internally to the timer (connector TST2).Jumpers are
    available to activate properly the hand-switch cPosition TST1 for
    TST1, TST2 for TST2, TST1/2 for both.

    3.1.12 Optional Remote Hand-Switch

    The hand-switch can be remotely mounted using the optionaremote
    box (wall holder) and connection cable.

    The wall holder (remote box) has a hook to hold the hanplug
    below where to connect the coiled cable.

    A cable 10 m long is provided with a plug on one side, tsocket
    on the timer, and free wires on the other side, to bconveniently
    around the room, and eventually to connectinto the wall holder
    (remote box).

    The connection of the wiresis here indicated.

    The wires of the switch arenumber 3 and 4 (those in the

    middle of the connector).

    GROUND

    TIMER CONNECTIO

    NEUTRAL IN

    LIVE IN

    INPUT

    FROMLINEVOLTAGE

    SUPPLY

    stallation Manual

    Version 4.0

    to

    0 V5x205x20A 5×20

    nations can be(connector

    nnection:

    kit made of a

    -switch and a

    connect to thee able to place itit to the block

    S

    NEUTRAL OUT

    LIVE OUT

    OUTPUT

    TOFOLDINGARM

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    21/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    3.1.13 Mounting and Connecting the Tube-Head

    A Remove retaining screws and cover plate on inner
    side(handgrip) of Folding/Simple Arm at tube-head side.

    B Unscrew and slide out the U shaped retaining clip .

    C Slightly lubricate the outer pin of the connecting sockeD
    Slide the connecting socket into the terminal pole of th

    Arm.E Put back the U shaped retaining clip making sure it
    has

    specific slot on the outer pin of the connecting socket alocking
    screw.

    F Put back the cover plate and block it with the four retai

    3.1.14 Mounting the Beam Limiting Device

    A Mount the beam limiting device on the tube-head and
    lclock-wise rotation.

    21/52

    of terminal pole

    of the tube-head.Folding/Simple

    entered thed tighten the

    ing screws.

    ck it close with a

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    22/52

    IntraOs 70 Service & In

    22/52 69 500 00210

    3.1.15 Final Tuning and Set-Up

    A Remove the line voltage fuse(s) from the timer

    B Put back the cover plate on the folding/simple arm (tubeC
    Tighten locking screw and tune friction screw in ExtensiD Tune
    friction screw on the folding arm for rotation aroun

    axis.E Make sure the Timer switch is in the off position.F Set
    the dip switches SW2 as indicated in the following t

    a Enable/disable correction of exposure time to
    compefluctuations of line voltage.

    b Indicate the nominal line voltage of the tube-head (s

    c Set the pre-heating time according to type of X-ray itube-head
    assembly; the measured value is marked itube-head;

    G Put back the line voltagefuse(s).

    H Put back all plastic coversand logo strips.

    I Connect the system to theline voltage by switching the

    general line voltage switchon and/or plug the powercord into the
    wall socket.

    J Switch the timer on.K The system is now ready for

    Functional Check asdescribed in the Operators Manual.

    PRE-HEATING TIME SETTING — TYPICAL V

    Type of insert OCX/70-G RPre-Heating

    Time180 ms 1

    stallation Manual

    Version 4.0

    head side)n Arm.d the vertical

    define:nsate

    e label).

    sert used in thefront of the

    LUES

    8G070

    60 ms

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    23/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Enable/disable exposure timecorrection to compensatefluctuations
    of mains voltage.

    dip 1

    Setting Dose Compensation

    Disabled

    Enabled

    Indicate the nominal voltage ofthe tube-head (see label).

    dip 2 dip 3Setting Tube-Head Type

    110 or 220VAC

    115 or 230VAC

    120 or 240VAC

    125 or 250

    VAC

    Set pre-heatinindicated on ltube-head

    dip 5 dip 6

    Setting Filament Pr

    100 ms

    120 ms

    140 ms

    160 ms

    180 ms

    200 ms

    220 ms

    240 ms

    DIP 4NOT USED

    DIP 8NOT USE

    PRDIP 2-3

    TUBE-HEAD TYPEDIP 1

    DOSE COMPENSATION

    23/52

    time asbel in front of the

    dip 7

    -Heating Time

    DIP 5-6-7-HEATING TIMES

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    24/52

    IntraOs 70 Service & In

    24/52 69 500 00210

    3.1.16 Layout Power Board Timer AutoSet

    SET-UPMICROSWITCHES

    KEYBOARDCONNECTOR

    BACK UP TIMETEST POINT & LE

    MICROPROCESSORWATCH DOG LED

    FUSE 4T 200 mA

    5 x 20

    POSUP24 V

    FUSET 2005 x 2

    OUTPUTLINE TO

    TUBEHEAD

    INPUTMAINS LINENEUTRAL

    & LIVE

    JUMPERFUSE 2

    MAINSSWITCH

    FUSE 1& FUSE 2

    T8A 6.3×32

    stallation Manual

    Version 4.0

    POWERSUPPLY5 V LED

    D

    ERPLYLED

    3A

    HANDSWITCHSOCKETS

    ANDSWITCHJUMPERS

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    25/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    3.2 Mobile Systems

    3.2.1 Unpacking

    Unpack the components of the system andcheck the following:

    A Each item is in good conditions and wasnot damaged during
    transportation.

    B All the items for the desired systemconfiguration are
    available.

    C The line voltage on the labels of timerand tube-head are
    corresponding to theexisting local line voltage.

    3.2.2 Room Preparation

    Make sure that a wall socket is availableclose to the dental
    chair in reach of themobile X-ray equipment to be connected.

    Such a socket has to be provided with aground connector to the
    protective earthpoint of the room. For the line voltage

    supply to the wall socket it is recommendedthe use of a cable
    with large sectionconductors to keep to minimum the cableresistance
    and the relevant voltage drop.

    Wire Type Square Section mm ResiAWG 16 1.35AWG 14 2.11

    25/52

    tance Ohm/m2x0.015×0.0094

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    26/52

    IntraOs 70 Service & In

    26/52 69 500 00210

    3.2.3 Mounting and Connecting Sequence

    The recommended sequence to mount andconnect system modules is
    listed here below.

    A Assembly the Mobile Stand.B Mount the Timer.C Mount the
    Suspension Arm.D Connect the Line voltage Cable.E Connect the
    Timer.F Mount and connect the Tube-head.G Mount the Beam Limiting
    Device.H Final Tuning and Set-Up.

    3.2.4 Assembling the Mobile StandThe following sequence to be
    followed.

    A Put together the base elements and tightenscrew and
    washer.

    B Mount the column (timer plate externally tothe long legs see
    picture) and tighten itwith the screw and washer from below.

    C Mount the front wheels and rear wheels

    with brake.D Mount the manoeuvring handles.E Connect the plug on
    the line voltage cable.

    3.2.5 Mounting the Timer

    A Refer to section 0 at page 18.

    B The zero Ohm resistor R51

    close to fuse F2 has to be cutfor systems cord connected tothe
    mains.

    3.2.6 Mounting the Suspension Arm

    Refer to section 3.1.8 at page 17.

    stallation Manual

    Version 4.0

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    27/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    LN

    INPUTFROMLINEVOLTAGESUPPLY

    GROUPOINT

    GR

    GROUNDWIRE INPUT

    NEUTRAL IN

    TIMER CONNECT

    LIVE IN

    3.2.7 Connecting the Line voltage Cable

    A The line voltage cable for the mobile unit is providedB The
    grounding wire of the main cable has to be blocke

    grounding provision (common bolt for bonding and gr

    nut and lock-washer.C The grounding wire for the timer is the
    placed on the c

    bonding and grounding and blocked (second) with nutD Connect the
    grounding cable for the timer to

    the ground provision on the terminal block ofthe timer and the
    two linewires from the mains cable tothe input point for live
    andneutral connections.

    3.2.8 Connecting the Timer

    For the connection of the outputcable to the timer please refer
    tosection 0 at page 18.

    3.2.9 Activation ofFuse 2

    Units provided with supply cordand plug must be equipped with
    fuses onboth line phases. The activation on thesecond fuse is done
    by cutting the JumperFuse 2 (zero Ohm resistor) close to FuseF2.
    See position on layout in section3.1.16 at page 24.

    3.2.10 Connecting the Tube-HeadRefer to section 3.1.11 at page
    20.

    3.2.11 Mounting the Beam Limiting Device

    Refer to section 3.1.14 at page 21.

    3.2.12 Final Tuning and Set-Up

    Refer to section 3.1.15 at page 22.

    27/52

    LN

    OUTPUTTO

    FOLDINGARM

    DS

    OUND PROVISIONN MOBILE FRAME

    IONS MOBILE

    ith plug.(first) to the

    unding) using

    mmon bolt fornd lock-washer.

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    28/52

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    29/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    FRICTIOSCRE

    4.1 Maintenance of the WallSupport

    A Remove the cover and verify that the mountingis closely
    connected to the wall and stays firmand steady during various
    movements of thesystem.

    B Verify that the ground cable of the wall supportis properly
    connected to the wall plate and tothe blocking bolt on the
    extension arm.

    C Check friction during rotation of ExtensionArm and adjust
    relevant screw if needed.

    D Verify that the technical label withidentification data
    outside the plastic cover atbottom is in place and is readable.

    4.2 Maintenance of the Extension

    A Slide partially out

    the extension armfrom wall mount,inspect for wear ofthe joint
    andlubricate for smoothrotation.

    B Check frictionduring rotation ofFolding Arm andadjust relevant
    screwif needed.

    C Verify that thetechnical label withidentification data isin
    place and isreadable.

    29/52

    N

    rm

    FRICTIONSCREW

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    30/52

    IntraOs 70 Service & In

    30/52 69 500 00210

    ARM FIRSTSECTION

    MINIMUMCLEARANCE

    ARM SECONDSECTION

    a) BASIC

    b) BRAS

    c) SAFE

    d) SIDE

    PARTS OF ARTICULATED JOI

    a

    b

    cd

    4.3 Maintenance of the Folding ArmSpecifically for the folding
    arm inspect for wear of pins and

    every 24 months, in case of normal use (about 6000 cycles).

    worn out parts or the arm itself if it appears damaged.

    Maintenance can be performed with the arm mounted on theor on
    the mobile stand.

    A Disconnect the tube housing assembly keeping the arm favoid
    the risk of injuries (fig.1) .

    B Remove the snap-out caps of the first arm section, the blof
    the pins, then the metal cover by pushing the pins outattention to
    supporting the arm (fig. 2 and fig. 3).

    stallation Manual

    Version 4.0

    PIN

    BUSHING

    Y CLAMP

    OLDER

    T

    evers at least

    eplace the

    extension arm

    lly open to

    cking screwsaying

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    31/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Fig. 4

    Fig. 1 Fig. 2

    Fig. 5

    Fig. 8

    BENT ARM

    (IN SHADE)INDICATESWEARING OUTOF BASIC PIN

    31/52

    Fig. 9

    Fig. 6

    Fig. 7

    Fig. 3

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    32/52

    IntraOs 70 Service & In

    32/52 69 500 00210

    Fig. 10

    Fig. 13

    Fig. 15

    Fig. 17

    Fig. 11

    stallation Manual

    Version 4.0

    Fig. 12

    Fig. 14

    Fig. 16

    Fig. 18

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    33/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    C Be careful not to pinch fingers between the two sectionwhile
    manoeuvring (fig. 4).

    D In the second section of the arm remove the two
    snap-otube-head side and the pin in order to raise the metal c

    friction (fig. 10) and spring (fig. 11). In case of removathe
    metal cover of the second section is different.E Inspect the main
    joint at the base of the arm (fig. 8) for

    pin and levers. A clearance lower than 3 mm of the blotwo levers
    from the body of the firstarm (fig. 7) indicates that the basicpin
    of the articulated joint iswearing out, and must be replaced.Also a
    parking position with the firstsection of the arm not
    standingvertical (fig. 4) but bent to some 5degrees or more reveals
    that the basicpin is wearing out. To replace the pinof the
    articulated joint (fig. 8)proceed as follows:a Loosen the friction
    screw (fig. 10) and release the s

    rotating the spring cap (fig. 11) with a round tool o(the base
    of a drill is fine) until zero pressure on th12, about 1 cm free
    thread left);

    b Keeping the first arm vertical, push the brass bushiand
    extract the supporting parts (round tool size 10

    c Disassemble the parts of the joint and replace the bd Tension
    the spring as needed by screwing the sprin

    desired distance (fig. 11, about 5 cm free thread lefF Lubricate
    moving parts (fig. 17 and fig. 18) and the thr

    the metal covers before their insertion.G Mount the metal covers
    provisionally, with friction scr

    H Mount the tube-head and secure it with the holding forI Verify
    proper balancing and in case tune tension of sprJ Tune the friction
    screws of the two arms.K Block tight the screws of metal covers and
    place the snL Slide partially out the Arm from the Extension Arm,
    in

    the vertical joint and lubricate for smooth rotation.M Verify
    that the label with identification data is in place

    CLEC

    33/52

    s of the arm

    ut caps on thever and tune

    l, beware that

    wearing out ofk holding the

    pring by4 mm diameterbasic pin (fig.

    g out (fig. 13)mm diameter);asic pin;g cap to the);e pins to
    block

    ws loosen

    (fig. 1)ngs.

    p-out caps.pect for wear of

    and is readable.

    3 mmMINIMUM

    RANCEECK

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    34/52

    IntraOs 70 Service & In

    34/52 69 500 00210

    FRICTIONSCREW

    FRICTIONSCREW

    4.4 Maintenance of the Tube-HeadA Inspect for damage/wear to the
    X-ray tube-head and supB Inspect for oil leakage. Replace tube-head
    if necessary.C Check for position

    stability. Adjustfriction for rotationaround the horizontalaxis
    if necessary.

    D Check that electricalcables are correctlytwisted and
    notdamaged or wear out.

    E Verify that thetechnical label withidentification data isin
    place and isreadable.

    F Dismount the tube-head from the arm and lubricate the pin
    forsmooth rotation. Warning: remove the tube-head with the arm
    completely open to avoidthe risk of injuries.

    G Check the U shaped retaining clip and replaceif wear. Do not
    mount the tube-head until allchecks of the suspension system
    arecompleted.

    4.5 Maintenance of the Beam LimitiA Verify that the collimator
    is properly mounted and, in ca

    firmly closed.

    BVerify that the internal cone is not damaged. Replace
    thenecessary.

    C Verify that the technical label with identification data
    isreadable.

    4.6 Maintenance of the TimerA Inspect the panel for damages on
    the surface.B Inspect the cable of the exposure switch for wear.C
    Check that all lights (except the yellow one) are working

    system is switched on.D Perform a Functional Check as described
    in the Operator

    This check includes verification of: a) yellow light, b) bu

    stallation Manual

    Version 4.0

    FRICTIONSCREW

    ort system.

    g Devicee, lock it

    collimator if

    n place and is

    when the

    s Manual.zzer, c) alarm

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    35/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    for Exposure push button pressed at power on, d) alaExposure
    stopped by the operator.

    E Verify that the fuses mounted comply with what
    specifbelow.

    F Verify that the technical label with identification data
    ireadable.

    Table of Fuses for Timer AutoSetFuse 115 V 2

    F1 T8 A 6.3×32 T5F2 T8 A 6.3×32 T5

    F3, F4 T200 mA 5×20 T200

    4.7 Checking Filament Pre-HeatingDuring the life of the system
    thewearing out of the filament may resultin a different pre-heating
    time (PHT).The value set at installation needstherefore to be
    verified and in casecorrected.A At a fixed distance measure the
    X-

    ray dose for exposure of 2.0 s

    (nominal irradiation of 100 pulsesat 50 Hz, 120 at 60 Hz).

    B At same distance measure the dose for exposure of 0.2(nominal
    irradiation of 10 pulses at 50 Hz, 12 at 60 Hz

    C If dose at 0.2 s is 1/10 of dose at 2 s, the value set for PD
    If dose at 0.2 s is lower than 1/10 of the dose at 2 s, PH

    increased. Proceed at steps of 20 ms until fine.E If dose at 0.2
    s is higher than 1/10 of the dose at 2 s, P

    reduced. Proceed at steps of 20 ms until fine.

    4.8 Special ServiceFunctions

    A Digital Volt Meter. This function allows theuser to display
    the line voltage level (forexample 224 V in the side picture).a
    Enter this function by switching the system

    on while pressing at the same time the

    , the , and the keys.b Exit by switching the system off.

    35/52

    m for

    ed in the table

    in place and is

    30 V5x205x20

    mA 5×20

    Time

    s.

    T is fine.T has to be

    T has to be

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    36/52

    IntraOs 70 Service & In

    36/52 69 500 00210

    B Test of Control Panel. This function is intended to test liand
    segments of numerical display on the control panel.a Enter this
    function by switching the system ON w

    the same time the , the , and the adult keys.

    b By pressing the or keys all segments of the ndisplay are
    scanned.c By pressing the X-ray pushbutton the rAY mes

    displayed.d By pressing each button on the panel, the
    correspon

    turned ON; a second pressure would turn it OFFe At the last
    button pressed the system automatically e

    function.WARNING: Do not press the X-ray pushbutton as l

    button twice; the second time the timer becomes actiand
    radiation is emitted.

    5. MEASUREMENTS

    All measurements to be done with the equipment supplied
    atvoltage with specified line resistance (see section 2.2 at
    pagetolerances on the measured values have to take into account
    tthe each measurement instrument. Measurements have to be

    trained personnel only to avoid the risk of electrical
    shock.

    5.1 Line VoltageThe Line voltage can be measured using a Volt
    Meter for Altwithin the proper range. A service function can be
    activatedtimer to display the current line voltage (see section 4.8
    of th

    5.2 Anode Voltage KVpThe kVp level is the actual peak value of
    the Anode voltage

    stabilizes once the preheating time of the filament has
    elapse0.2 s) and the high voltage transformer is actually loaded.
    Thcan be measured with a non-invasive kVp meter placed in
    frfollowing the instructions in the user manual of the
    instrumemeasure can be done with an exposure time of 500 ms or
    mointroduced a delay of the reading instrument of about 300
    msvoltage level to stabilize after the pre-heating time has
    elapse

    5.3 Anode Current mA

    WARNING. To prevent high voltage shock make sure the syis
    disconnected from the power supply when connections to tmeasurement
    points are performed. Electric discharge might

    stallation Manual

    Version 4.0

    hts, buttons,

    ile pressing at

    merical

    sage is

    ing lamp is.xit this

    st

    e

    nominal line9). Thehe precision ofdone by

    ernate Currentn the AutoSet

    is Manual).

    hich

    (in aboute kVp levelnt of the BLDt. A correcte havingto allow
    thed.

    stemhe

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    37/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    A1

    A2

    occur in case of improper operation

    The Anode current in mA is the actualaverage value of the tube
    current whichraises when the filament has warmed-

    up, after the preheating time haselapsed.

    Remove the back cover of the tube-head, disconnect the jumper
    A1-A2 andreplace it with a precision resistor of1000 Ohm.

    The anode current can be read with aDC voltmeter connected to
    points A1 and A2, with the 1 Vto 1 mA.

    After the measurement disconnect the 1000 Ohm resistor
    ajumper.

    5.4 Exposure TimeThe actual exposure (loading) time is
    determined as the tibetween the instant when the tube current first
    rises above tmaximum value and the instant when it finally falls
    below

    In order to assure the requested exposure time
    (irradiation),

    time of the used insert, set-up at installation, is
    automaticalconsideration by the microprocessor.

    The actual switch-on time (SOT) of the tube-head assemblsum of
    the filament pre-heating time (PHT) and of the requtime (RET).

    SOT = PHT + RET

    During filament pre-heating radiation is not emitted.

    The measure of the exposure time can be correctly measureTime
    meters with the ability to trigger when the kV level rmaximum.

    Exposure time can also be measured after having connecteresistor
    of 1000 Ohm between points A1 and A2 on the smon the back of the
    tube-head, by monitoring the voltage acrcurrent) with an
    oscilloscope, and counting the number ofinterval, as per above
    definition.

    When the compensation function to correct the effects of lion
    the emitted X-ray dose is activated, the requested expos

    corrected in order to assure dose consistency.When the actual
    line voltage is lower than the nominal lineactual exposure time is
    increased.

    37/52

    corresponding

    d put back the

    e intervalhe 25% of itshe same value.

    the pre-heating

    y taken into

    is therefore theested exposure

    d with kVp &aches 75% of its

    a precisionll circuit board

    oss it (tubeulses in the time

    e fluctuationsre time is

    voltage, the

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    38/52

    IntraOs 70 Service & In

    38/52 69 500 00210

    When the actual line voltage is higher than the nominal
    lineactual exposure time is decreased.

    Do disable the compensation function (dip switch number 1to OFF
    position) to verify exposure time accuracy.

    This is not necessary in case a meter with the ability to
    triggthe kVp level and the value read should be compared with
    thgiven by the timer for 3 seconds after the exposure (refer to
    tManual to activate the function to display the actual
    exposurcompensation).

    5.5 Leakage RadiationThe value of leakage radiation of the X-ray
    tube assembly is1 m distance taking into account the energy
    provided in 1 ho

    measurement is performed at a distance different from 1 m, tbe
    properly corrected.

    The applicable technique factors are of 70 kV and 7 mA.
    Thmaximum leakage has to be considered, excluding the
    directiprimary beam. The duty cycle factor as reported in
    sectionsbe taken into account for the computation of the actual
    dose

    In case for example a measure is done at 50 cm distance frowith
    an exposure of 1 s, the following computations have to

    A duty cycle of 1/30 means maximum of 1 s exposure e120 s per
    hour.

    Dose at 100 cm is computed taking into account the invei.e. at
    twice the distance the dose intensity is reduced toD1h100cm =
    D1s50cm * 120 * (50/100)2 = D1s50cm *where: D1s50cm is the dose
    measured at 50 cm with 1 sand D1h100cm is the dose computed at 1 m
    distance forhour.

    5.6 Earth ResistanceThe measurement of the earth resistance to
    be done after havdisconnected the system for the power supply
    line.

    Using a voltage generator of not more than 6 V, 50/60 Hz, ais
    applied from 5 s to 10 s through the earth terminal to anywhich can
    be in touch to the patient/operator.

    The voltage drop is then measured and the earth resistance c

    The system complies for values lower than 0,1 (extended t

    the supply cable cannot be disconnected like in the cases
    ofinstallations).

    stallation Manual

    Version 4.0

    oltage, the

    f SW2 moved

    r on 75% ofindication

    he Operatorstime after

    measured atr. In case the

    e value has to

    point ofon of the.2 has also toate.

    the sourcee performed:

    ery 30 s, i.e.

    se square law,ne quarter.0

    exposurethe energy of 1

    ing

    urrent of 25 Aetal part

    mputed.

    o 0,2 in case

    ermanent

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    39/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    5.7 Earth LeakageConnect the Timer to themeasuring circuit
    asshown in the figure(R1=10 k5%, R2=1k1%,C1=0,015 F 5%)

    A Set the input supplyvoltage (Vp) equalto 110% of thenominal
    supplyvoltage.

    B Connect a voltmeter(Vx with 1 MOhminput impedance) tothe
    indicated points

    C Switch the Timer OND Make two measures in Normal Conditions
    (NC with S1

    S5 ON: live and neutral conductors in normal co

    S5 OFF: swap of live and neutral conductors.

    E Make two measures in Single Fault Condition (SFC wi S5 ON:
    live and neutral conductors in normal co

    S5 OFF: swap of live and neutral conductors.

    Please note that 1V equals to 1mA.

    Make sure values of leakage currents fall within the limitstable
    below.

    LIMIT OF EARTH LEAKAGE CURREN

    Type Of Device

    Normal

    Condition(S1 ON)X-ray Equipment permanentlyconnected to line
    voltage 5 mAMobile X-ray Equipment 2,5 mA

    39/52

    ON)dition.

    h S1 OFF).dition.

    hown in the

    Tingle Fault

    ConditionS1 OFF).

    10 mA5 mA

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    40/52

    IntraOs 70 Service & In

    40/52 69 500 00210

    6. Spare Parts

    6.1 Tube-Head

    93 200 01700: G 230 V without BLD93 200 01300: G 120 V without
    BLD

    List of Spare Parts Figure THItem Description

    A Rear and Front Cover

    BRound Beam Limiting Device, FFD 21 cmOutput Field 6 cm
    diameter

    C rectangular Beam Limiting Device, FFD 21 cmOutput Field
    3.5×4.6 cm

    D Side cap tube-head holder

    ECoupling with Concentric Pin

    for SN lower than TK0501F Friction Ring for tube-head with
    screwsG Wave Suppressor

    Figure TH

    A

    B

    D

    F

    C

    stallation Manual

    Version 4.0

    Code76 190 25110

    91 300 00020

    91 300 00040

    76 190 25120

    76 190 2524176 190 2524076 190 2545076 190 25600

    E

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    41/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    6.2 Folding Arm

    93 100 12010: for Wall Mount

    List of Spare Parts — Figure FAItem Description

    ATube head support for folding arm

    for SN lower than FA0501B Cover Plate Tube-Head HolderC Arm
    Cover Extension SideD Arm Cover Tube-Head SideE Retaining Clip With
    Screw

    FCable 2.8 m With Concentric Connector

    for SN lower than FA0501G Ring Spacer D 25 With HoleH Snap Out
    Cap (set of 8)I Friction Screw M6x10 For Arm (set of 2)L Basic pin
    and seeger rings

    Figure FA

    AD

    C

    H

    F

    G

    I

    L

    41/52

    Code76 190 2525176 190 2525076 190 2513076 190 2514076 190
    2515076 190 2527076 190 2543176 190 2543076 190 2546076 190 2516076
    190 2534076 500 30010

    B

    E

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    42/52

    IntraOs 70 Service & In

    42/52 69 500 00210

    6.3 Extension Arm93 100 17100 45 cm93 100 17200 70 cm93 100
    17300 90 cm

    93 100 17400 110 cm

    List of Spare Parts Figure EAItem Description

    A Large Ring Spacer D35B Snap out Cap D25 (set of 8)

    CFriction Screw Brass Cap M6x10 (set of 3)

    Friction Screw Brass Cap M8x10 (set of 3)

    Figure EA

    A B

    stallation Manual

    Version 4.0

    Code76 190 2547076 190 2518076 190 2533176 190 25332

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    43/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    6.4 Wall Support93 100 11000

    List of Spare Parts — Figure WSItem Description

    A Plastic Cover Wall MountB Logo Strip Blue X

    Figure WS

    A

    B

    43/52

    Code76 190 2519076 190 25200

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    44/52

    IntraOs 70 Service & In

    44/52 69 500 00210

    6.5 Mobile Base93 100 2008093 100 20090

    List of Spare Parts Figure MBItem Description

    A Transport Handle (set of 2)B Cable Holder With Strain ReliefC
    Hook For CableD Wheel (set of 2)E Wheel With Brake (set of 2)

    Figure MB

    A

    B

    C

    DE

    stallation Manual

    Version 4.0

    Code76 190 2528076 190 2530076 190 2529076 190 2541076 190
    25420

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    45/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    6.6 AutoSet Timer93 300 60100: 230 V93 300 60200: 115 V

    List of Spare Parts Figure AUItem Description

    A Front CoverB Mounting PlateC Mounting Columns

    D230V Power Control Board With Triac115V Power Control Board
    With Triac

    E Keyboard Control PanelF AutoSet Membrane

    G Hand Switch Without CableH Coiled CableI Logo Strip IntraOs
    70

    J1230V Fuse F1/F2 T5A 250V 5×20 (set of 10)115V Fuse F1/F2 T8A
    250V 6.3×32 (set of 10)

    J2 Fuse F3/F4 T200mA 250 V 5×20 (set of 10)

    Figure AU

    A

    BE

    C

    HG

    I

    45/52

    Code76 190 2522076 190 2523076 190 2531076 190 2551076 190
    2551176 190 2553076 190 25550

    76 190 2559076 190 2558076 190 2521076 190 2564076 190 2564176
    190 25630

    D

    F

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    46/52

    IntraOs 70 Service & In

    46/52 69 500 00210

    Appendix ASystem Components

    ArticleSystem Component

    Type CodeCatalogue Number

    Wall Support 93 100 11000

    Extension Arm 45 cm 93 100 17100Extension Arm 70 cm 93 100
    17200Extension Arm 90 cm 93 100 17300

    Extension Arm 110 cm 93 100 17400

    Folding Arm 93 100 12010

    Tube Head G 120 V 93 200 01300Tube Head G 230 V 93 200 01700

    20 cm Round BLD 91 300 0002020 cm Rectangular BLD 91 300
    00040

    AutoSet Timer 115 VAC 93 300 60200AutoSet Timer 230 VAC 93 300
    60100

    Mobile Base 93 100 20080Mobile Base UL/CSA 93 100 20090

    Wall Plate 86 100 11500

    stallation Manual

    Version 4.0

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    47/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Appendix BIcons

    IEC Type BEquipment

    ComplComm

    X-ray OnComplUS St

    Examine AnnexedDocumentation

    Line vSyste

    Increase Exposure Time(one step)

    Off (Divoltag

    Decrease Exposure Time(one step) On (Cvoltag

    Child Small Patient ~ Altern

    Adult Large Patient Fuse

    Upper Incisor Protec

    Upper Canine/Premolar N Neutraperma

    Upper Molar LLive Pperma

    Lower Incisor Inhere

    Lower Canine/Premolar gggg Focal

    Lower Molar FragileHandl

    Bite Wing — Interproximal Fear o

    Digital ReceptorUpDo No

    Radiography Push Button Stacki

    4

    Ionizing Radiation SeparDo No

    47/52

    iance to Europeanunity Requirements

    iance to Canadian andndards

    ltage supply On -Ready

    sconnected from LineSupply)

    nnected to LineSupply)

    te Current

    ive Earth

    l Point (for equipmentent connected to line)

    int (for equipmentent connected to line)

    t Filtration

    pot

    ,With Care

    Humidity

    Overturn

    g Limit

    te Collection,Dispose

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    48/52

    IntraOs 70 Service & In

    48/52 69 500 00210

    Appendix CExposure Table

    IntraOs 70 — 70 kVp, 7 mA — Exposure Times in s

    Focus-F

    ilmD

    istance23cm

    DFilm

    EFilm

    DFilm

    EFilm

    Digital

    DFilm

    EFilm

    Digital

    3,20 2,50 2,00 2,60 1,25 1,00 0,80 0,64 0,50 0,40

    2,50 2,00 2,60 1,25 1,00 0,80 0,64 0,50 0,40 0,32

    2,00 2,60 1,25 1,00 0,80 0,64 0,50 0,40 0,32 0,25

    1,60 1,25 1,00 0,80 0,64 0,50 0,40 0,32 0,25 0,20

    LowerMolar

    1,25 1,00 0,80 0,64 0,50 0,40 0,32 0,25 0,20 0,16

    LowerPremolar

    1,00 0,80 0,64 0,50 0,40 0,32 0,25 0,20 0,16 0,12

    LowerIncisor

    0,80 0,64 0,50 0,40 0,32 0,25 0,20 0,16 0,12 0,10

    0,64 0,50 0,40 0,32 0,25 0,20 0,16 0,12 0,10 0,08

    F

    ocus-FilmD

    istance33cm

    DFilm

    EFilm

    Digital

    DFilm

    EFilm

    Digital

    D

    Film

    E

    Film

    Digital

    stallation Manual

    Version 4.0

    Digital

    SmallPatient

    LargePatient

    0,32

    0,25

    0,20UpperMolar

    0,16 UpperPremolar

    0,12 Bite Wing

    0,10UpperIncisor

    0,08

    0,06

    SmallPatient

    LargePatient

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    49/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Appendix DAlarm Conditions

    AutoSet Timer Alarm Conditions

    Code Fault /Error Signal Action Re

    A 01X-ray requestedduring cool-downperiod

    Green lamp(System Ready)flashing

    Systeminhibited

    Byaconsy

    A 02Line voltagebelow lower limit

    Green lamp (SystemReady) and red lamp(Alarm) flashing

    Systeminhibited

    Aulinra

    A 03 Line voltageabove upper limit

    Green lamp (SystemReady) and red lamp(Alarm) flashing

    Systeminhibited

    Aulinra

    A 04Correctedexposure timelower than 60 ms

    Green lamp (SystemReady) and red lamp(Alarm) flashing

    0.06 sforced.

    Byacon

    A 05Correctedexposure timegreater than 3.2 s

    Green lamp (SystemReady) and red lamp(Alarm) flashing

    3.2 sforced.

    Byacon

    A 06Line Frequency

    Detection Failure

    System Ready(green) lamp and

    Alarm (red) lampflashing

    System

    inhibited

    By

    off

    A 07Exposure pushbutton pressed atpower on

    Red lamp(Alarm) flashing

    Exposureinhibited.

    Byacon

    A 08Exposurestopped by theoperator

    Red lamp(Alarm) flashing

    Systeminhibited

    Byacon1

    A 09Exposurestopped by the

    back-up timer

    Red lamp (Alarm)switched on

    Systeminhibited

    Byoff

    A 10Back-up relayfailure

    Red lamp (Alarm)switched on

    Systeminhibited

    Byoff

    A 11Power switchingdevice failure

    Red lamp (Alarm)switched on

    Systeminhibited

    Byoff

    A12Line dips duringexposure

    Red lamp (Alarm)switched on

    Exposureinhibited

    Byacon

    49/52

    set

    nowledgementthe panel or whentem cooled down

    tomatically whenvoltage back in

    ge

    tomatically whenvoltage back inge

    nowledgementthe panel

    nowledgementthe panel

    switching system

    and on again

    nowledgementthe panel

    nowledgementthe panel or after

    switching systemand on again

    switching systemand on againswitching systemand on again

    nowledgementthe panel

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    50/52

    IntraOs 70 Service & In

    50/52 69 500 00210

    Appendix EIdentification Labels

    stallation Manual

    Version 4.0

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    51/52

    IntraOs 70 Service & Installation Manual

    Version 4.0 69 500 00210

    Appendix FCooling Curves

    51/52

  • 8/10/2019 IntraOs 70 V4.0 ENG — Service Installation

    52/52

    IntraOs 70Dental X-ray Equipment

    Service & Installation ManualEnglish Edition

    Version 4.0

Code Fault /Error

A 01

X-ray requested

during cool-down

period

A 02

Line voltage

below lower limit

A 03

Line voltage

above upper limit

A 04

Corrected

exposure time

lower than 60 ms

A 05

Corrected

exposure time

greater than 3.2 s

A 06

Line Frequency

Detection Failure

A 07

Exposure push

button pressed at

power on

A 08

Exposure stopped

by the operator

A 09

Exposure stopped

by the back-up

timer

A 10

Back-up relay

failure

A 11

Power switching

device failure

A12

Line dips during

exposure

Appendix D

Alarm Conditions

AutoSet Timer Alarm Conditions

Signal

Green lamp

(System Ready)

flashing

Green lamp (System

Ready) and red lamp

(Alarm) flashing

Green lamp (System

Ready) and red lamp

(Alarm) flashing

Green lamp (System

Ready) and red lamp

(Alarm) flashing

Green lamp (System

Ready) and red lamp

(Alarm) flashing

System Ready (green)

lamp and Alarm (red)

lamp flashing

Red lamp

(Alarm) flashing

Red lamp

(Alarm) flashing

Red lamp (Alarm)

switched on

Red lamp (Alarm)

switched on

Red lamp (Alarm)

switched on

Red lamp (Alarm)

switched on

IntraOs 70 – Operator’s Manual

Action

System

inhibited

System

inhibited

System

inhibited

0.06 s

forced.

3.2 s

forced.

System

inhibited

Exposure

inhibited.

System

inhibited

System

inhibited

System

inhibited

System

inhibited

Exposure

inhibited

Reset

By acknowledgement

on the panel or when

system cooled down

Automatically when

line voltage back in

range

Automatically when

line voltage back in

range

By acknowledgement

on the panel

By acknowledgement

on the panel

By switching system

off and on again

By acknowledgement

on the panel

By acknowledgement

on the panel or after

1 m

By switching system

off and on again

By switching system

off and on again

By switching system

off and on again

By acknowledgement

on the panel

Ошибка А02 в рентгене FONA XDG

Sead1995 Дата: Четверг, 20.Окт.2016, 13:16 | Сообщение # 1

Стажер

инженер компании Harwind

OFFлайн

Украина



Киев, Украина

Добрый день.
Был установлен новый рентген FONA XDG. После недельной работы снимки начали светлеть, а на экране появилось сообщение «А02». В инструкции данная ошибка значит, что слишком малое напряжение поступает на рентген. Но при измерении на тестере выдаёт 219 В. При установке стабилизатора на 1.5 КВт ошибка начала то появляться, то исчезать. Снимки также засветвлённые.
Может кто–то сталкивался с такой проблемой, подскажите, пожалуйста, решение. Резистор в рентгеновской головке целый, плата тоже.
И подскажите какие предохранители и стабилизаторы нужно устанавливать для рентгенов.

 
madmac Дата: Суббота, 22.Окт.2016, 12:56 | Сообщение # 2

стоматология и связанное с ней

OFFлайн

Российская Федерация



СПб

ошибка во время съемки или в любое время? контроль сети надо проводить сутки-двое-неделю. и не тестером. есть приборы контроля качества электричества с памятью, подключил в соседнюю розетку, грубо говоря, а через пару дней снял данные. часто очень интересные вещи читаются…
линия от щитка на рентген отдельная или ещё что-то висит? компрессор, автоклав…
сколько фаз заведено в клинику? сколько чего на каждой фазе висит?
рентген сильно похож на блюикс 70. на плате сверхе есть ряд переключателей 8 штук? документация осталась? проверить установки по напряжению- с завода могло и на 230, и на 240 придти. тогда при просадке до 210- ошибка.

Сообщение отредактировал madmacСуббота, 22.Окт.2016, 12:57

 
Anakonda2a Дата: Суббота, 22.Окт.2016, 13:10 | Сообщение # 3

Техник

Инженер по Т.О. мед. техники

OFFлайн

Российская Федерация



Подольск

Цитата Sead1995 ()

Был установлен новый рентген FONA XDG

А вызвать спецов по гарантии не пробовали??

 
Sead1995 Дата: Суббота, 22.Окт.2016, 15:00 | Сообщение # 4

Стажер

инженер компании Harwind

OFFлайн

Украина



Киев, Украина

Цитата madmac ()

ошибка во время съемки или в любое время? контроль сети надо проводить сутки-двое-неделю. и не тестером. есть приборы контроля качества электричества с памятью, подключил в соседнюю розетку, грубо говоря, а через пару дней снял данные. часто очень интересные вещи читаются…
линия от щитка на рентген отдельная или ещё что-то висит? компрессор, автоклав…
сколько фаз заведено в клинику? сколько чего на каждой фазе висит?
рентген сильно похож на блюикс 70. на плате сверхе есть ряд переключателей 8 штук? документация осталась? проверить установки по напряжению- с завода могло и на 230, и на 240 придти. тогда при просадке до 210- ошибка.

Ошибка происходит в любое время, когда напряжение опускается ниже 218 В (по показаниям входного напряжения на стабилизаторе). Когда напряжение выше — ошибка исчезает и можно делать снимки. Но доктор жаловался, что с самого начала снимки очень светлые были, а сейчас вообще полностью белые. Скорее всего это связано с нехваткой анпряжения для разогрева рентгеновской трубки.
А можете посоветовать какой лучше приобрести прибор контроля качества электричества с памятью. Очень пригодился бы. А то так показания можно снималь или с тестера, или с экранов стабилизатора.
Сначала линия общая была и он поработал буквально неделю. Сейчас провели отдельную линию с щитка — и тот-же результат.
По поводу количества фаз в клинике и сколько чего на каждой фазе — это нужно у местных электриков узнавать. Знаю только, что на фазе, на которой изначально работал рентген, после клиники шли целый подьезд квартир (клиника расположена на первом этаже пятиэтажки).
Да, на плате рентгена вверху есть ряд из 8 микропереключателей и судя по инструкции на них установлено 220 В переменного тока (минимальное значение). Но сами показания входного напряжения на стабилизаторе редко превышают 220 В (особенно вечером).
Хоть бери и отдельный генератор ставь)

Сообщение отредактировал Sead1995Суббота, 22.Окт.2016, 15:16

 
Sead1995 Дата: Суббота, 22.Окт.2016, 15:03 | Сообщение # 5

Стажер

инженер компании Harwind

OFFлайн

Украина



Киев, Украина

Цитата Anakonda2a ()

А вызвать спецов по гарантии не пробовали??

Специ по гарантии аж в Словакии находятся и не поедут в Киев(

 
madmac Дата: Суббота, 22.Окт.2016, 16:39 | Сообщение # 6

стоматология и связанное с ней

OFFлайн

Российская Федерация



СПб

есть дилеры в Киеве и Запорожье, которые, по идее, должны поддерживать гарантию. а как по факту…
уже были стабилизаторы, которые только пишут, что на выходе 220, а так что на входе, то и на выходе.
кроме того, аппарат при съемке просаживает сеть, если длина кабеля на него большая и сечение не то.
опять же в жилых домах утро и вечер- пик потребления жильцами.
а прибор- проще поискать, где в прокат дают. покупать весьма небюджетно.

 
Sead1995 Дата: Суббота, 22.Окт.2016, 18:54 | Сообщение # 7

Стажер

инженер компании Harwind

OFFлайн

Украина



Киев, Украина

Цитата madmac ()

есть дилеры в Киеве и Запорожье, которые, по идее, должны поддерживать гарантию. а как по факту…
уже были стабилизаторы, которые только пишут, что на выходе 220, а так что на входе, то и на выходе.
кроме того, аппарат при съемке просаживает сеть, если длина кабеля на него большая и сечение не то.
опять же в жилых домах утро и вечер- пик потребления жильцами.
а прибор- проще поискать, где в прокат дают. покупать весьма небюджетно.

Я по сути и работаю в компании, которая продает и обслуживает данную продукцию. Но так как полноценного обучения я еще неуспел пройти, то учусь по ходу дела)
Да да, и в этой клинике именно такой стабилизатор и стоит: что на входе — то же и на выходе. И вот такая беда с сетью, даже непонятно что делать.
Ааа, понимаю, слишком дорогие. Будем искать прокат.
Спасибо вам за советы.

 
madmac Дата: Суббота, 22.Окт.2016, 20:12 | Сообщение # 8

стоматология и связанное с ней

OFFлайн

Российская Федерация



СПб

Цитата Sead1995 ()

по сути и работаю в компании, которая продает и обслуживает данную продукцию.

тогда не худо бы купить. пригодится и для рентгенов, и компрессора с автоклавами любят хорошее питание. а контроль перед монтажом нужен.
в старых городах, в центрах, очень часто такая ситуация- старые сети, старые подстанции. да ещё в жилых домах.
так-то надо брать за уши клиента и его электрика, это их задача -обеспечить нормальное ликстричество. но им надо ткнуть в нос параметры сети.

 
МастерБаку Дата: Воскресенье, 23.Окт.2016, 09:25 | Сообщение # 9

инженер

OFFлайн

Российская Федерация



Нижний Новгород

Цитата madmac ()

покупать весьма небюджетно.

nea
http://ekits.ru/index.php?productID=828
а то начинается — килянусь честный слово ничего не было .

вопрос в том — почему эта штука проводит экспозицию если напряжение мало ?

Сообщение отредактировал МастерБакуВоскресенье, 23.Окт.2016, 09:31

 
madmac Дата: Воскресенье, 23.Окт.2016, 12:56 | Сообщение # 10

стоматология и связанное с ней

OFFлайн

Российская Федерация



СПб

Цитата МастерБаку ()

http://ekits.ru/index.php?productID=828

не то малость. отображает максимум- минимум и все.
фиксация времени, периодичности изменений отсутствует, опять же чувствительность 6 вольт. если сидеть и втыкать в него сутки- пойдет. тогда и мультиметр вполне подходит. сиди с бумажкой в руках, см. на прибор и записывай.
вопрос в периодичности, времени, длительности провалов питания.

Цитата МастерБаку ()

вопрос в том — почему эта штука проводит экспозицию если напряжение мало ?

а пока в пределах+/- 10 вольт он и снимает.
Sead1995, ещё вопрос: время предпрогрева, установленное на плате, соответствует указанному на шильдике моноблока? и вообще замеры делали?

 
Sead1995 Дата: Воскресенье, 23.Окт.2016, 17:47 | Сообщение # 11

Стажер

инженер компании Harwind

OFFлайн

Украина



Киев, Украина

Цитата madmac ()

Sead1995, ещё вопрос: время предпрогрева, установленное на плате, соответствует указанному на шильдике моноблока? и вообще замеры делали

Замеры времени прогрева не делал. На заводе они сами устанавливают время прогрева для рентген трубки, которая в комплекте идет и остается только собрать. И сейчас выставленное время прогрева совпадает с тем, которое на моноблоке указано.
А как данные замеры производить?

 
grom1959 Дата: Четверг, 22.Мар.2018, 21:08 | Сообщение # 12

Заглянувший

сервисный инженер

OFFлайн

Российская Федерация



Спб

С FONA 70 такое же было, вылечилось покупкой 3 Кв стабилизатора.

 
Terracorp Дата: Среда, 29.Авг.2018, 08:46 | Сообщение # 13

Участник

Медтехник

OFFлайн

Российская Федерация



Уфа

Если я правильно понял. спалить аппарат переключением этих переключателей нельзя?
У меня на переключателях позиция, согласно фотографии, на 120 или 240 вольт. Если напряжение ниже 220 на 1 вольт, то загорается ошибка А02. Стабилизатор здесь не причем, так как переключает повышающую обмотку при 215 вольт, а при 216-219 вольт передает напрямую сетевое напряжение на потребители.

То есть у меня там никак не выйдет переключить на 120 вольт, вместо 220? Если исполнение на наши 220, то 110 вольт к нему не относится? Пардон, просто перестраховываюсь, мало ли, аппарат дорогой, отвечать деньгами нет желания.

Сообщение отредактировал TerracorpСреда, 29.Авг.2018, 08:48

 
Terracorp Дата: Вторник, 18.Сен.2018, 19:38 | Сообщение # 14

Участник

Медтехник

OFFлайн

Российская Федерация



Уфа

Вопрос решен, рентгеновский аппарат Fona X70 выдавал ошибка А02 при напряжении менее 220 вольт. Стабилизатором не лечилось, так как он в диапазоне от 216 до 225 работал в режиме байпас, и по стечению обстоятельств среднее напряжение в сети держалось на 216-219 вольт. Врач запускал рентген, включив чайник в подсобке, тогда срабатывала повышающая обмотка стабилизатора, что тоже не всегда помогало.

Итак, воспользовавшись фото схемы из сообщения №5, переключил программными переключателями (слева второй и третий) в режим работы 220 вольт (хотя до этого стояло на 230 — не работало, видно реально триггер срабатывал при 220, а ниже не хотел работать).

Врач на седьмом небе от счастья. Потому что сервисмены, которые продали и установили ему аппарат, тоже заметили глюки при включении, но предложили купить мощный стабилизатор, чтобы решить проблему. Как я уже писал, стабилизатор не помог, и врач отчаялся, навострив медсестру на манипуляции с чайником, светом и прочим, что выглядело весьма оригинально.

После этого не было ни одной ошибки А02. Если кто будет переключать, на всякий случай не забудьте делать это при обесточенном аппарате, так как я не проверял что будет, если переключать на включенном аппарате.

 
seabee Дата: Понедельник, 26.Июн.2023, 10:19 | Сообщение # 15

инженер

OFFлайн

Российская Федерация



Сибирь

Немного не в тему.
У нас Fona XDC. Трубка, наверно, такая же. Нет напряжения накала 4 вольта.
В моноблоке накал не прозванивается. Похоже простая схема-2 трансформатора и трубка.
Где можно заказать трубу SCAN-X OX/70-G4 7мА?

Сообщение отредактировал seabeeВторник, 27.Июн.2023, 05:36

 

Ошибка А02 в рентгене FONA XDG

Sead1995 Дата: Четверг, 20.Окт.2016, 13:16 | Сообщение # 1

Стажер

инженер компании Zdravo

На заявках

Украина

Киев, Украина

Добрый день. Был установлен новый рентген FONA XDG. После недельной работы снимки начали светлеть, а на экране появилось сообщение А02. В инструкции данная ошибка значит, что слишком малое напряжение поступает на рентген. Но при замере на тестере выдает 219 В. При установке стабилизатора на 1.5 КВт ошибка начала то появлятся, то осчезать. Снимки так–же засветвленные. Может кто–то сталкивался с такой проблемой, подскажите пожалуйста решение. Резистор в рентгеновской головке целый, плата тоже. И подскажите какие предохранители и стабилизаторы нужно устанавливать для рентгенов.
 
madmac Дата: Суббота, 22.Окт.2016, 12:56 | Сообщение # 2

Передовик

стоматология и связанное с ней

На заявках

Российская Федерация

СПб

ошибка во время съемки или в любое время? контроль сети надо проводить сутки-двое-неделю. и не тестером. есть приборы контроля качества электричества с памятью, подключил в соседнюю розетку, грубо говоря, а через пару дней снял данные. часто очень интересные вещи читаются…
линия от щитка на рентген отдельная или ещё что-то висит? компрессор, автоклав…
сколько фаз заведено в клинику? сколько чего на каждой фазе висит?
рентген сильно похож на блюикс 70. на плате сверхе есть ряд переключателей 8 штук? документация осталась? проверить установки по напряжению- с завода могло и на 230, и на 240 придти. тогда при просадке до 210- ошибка.

Сообщение отредактировал madmacСуббота, 22.Окт.2016, 12:57

 
Anakonda2a Дата: Суббота, 22.Окт.2016, 13:10 | Сообщение # 3

Техник

Инженер по Т.О. мед. техники

На заявках

Российская Федерация

Подольск

Цитата Sead1995 ()

Был установлен новый рентген FONA XDG

А вызвать спецов по гарантии не пробовали??

 
Sead1995 Дата: Суббота, 22.Окт.2016, 15:00 | Сообщение # 4

Стажер

инженер компании Zdravo

На заявках

Украина

Киев, Украина

Цитата madmac ()

ошибка во время съемки или в любое время? контроль сети надо проводить сутки-двое-неделю. и не тестером. есть приборы контроля качества электричества с памятью, подключил в соседнюю розетку, грубо говоря, а через пару дней снял данные. часто очень интересные вещи читаются…
линия от щитка на рентген отдельная или ещё что-то висит? компрессор, автоклав…
сколько фаз заведено в клинику? сколько чего на каждой фазе висит?
рентген сильно похож на блюикс 70. на плате сверхе есть ряд переключателей 8 штук? документация осталась? проверить установки по напряжению- с завода могло и на 230, и на 240 придти. тогда при просадке до 210- ошибка.

Ошибка происходит в любое время, когда напряжение опускается ниже 218 В (по показаниям входного напряжения на стабилизаторе). Когда напряжение выше — ошибка исчезает и можно делать снимки. Но доктор жаловался, что с самого начала снимки очень светлые были, а сейчас вообще полностью белые. Скорее всего это связано с нехваткой анпряжения для разогрева рентгеновской трубки.
А можете посоветовать какой лучше приобрести прибор контроля качества электричества с памятью. Очень пригодился бы. А то так показания можно снималь или с тестера, или с экранов стабилизатора.
Сначала линия общая была и он поработал буквально неделю. Сейчас провели отдельную линию с щитка — и тот-же результат.
По поводу количества фаз в клинике и сколько чего на каждой фазе — это нужно у местных электриков узнавать. Знаю только, что на фазе, на которой изначально работал рентген, после клиники шли целый подьезд квартир (клиника расположена на первом этаже пятиэтажки).
Да, на плате рентгена вверху есть ряд из 8 микропереключателей и судя по инструкции на них установлено 220 В переменного тока (минимальное значение). Но сами показания входного напряжения на стабилизаторе редко превышают 220 В (особенно вечером).
Хоть бери и отдельный генератор ставь)

Сообщение отредактировал Sead1995Суббота, 22.Окт.2016, 15:16

 
Sead1995 Дата: Суббота, 22.Окт.2016, 15:03 | Сообщение # 5

Стажер

инженер компании Zdravo

На заявках

Украина

Киев, Украина

Цитата Anakonda2a ()

А вызвать спецов по гарантии не пробовали??

Специ по гарантии аж в Словакии находятся и не поедут в Киев(

 
madmac Дата: Суббота, 22.Окт.2016, 16:39 | Сообщение # 6

Передовик

стоматология и связанное с ней

На заявках

Российская Федерация

СПб

есть дилеры в Киеве и Запорожье, которые, по идее, должны поддерживать гарантию. а как по факту…
уже были стабилизаторы, которые только пишут, что на выходе 220, а так что на входе, то и на выходе.
кроме того, аппарат при съемке просаживает сеть, если длина кабеля на него большая и сечение не то.
опять же в жилых домах утро и вечер- пик потребления жильцами.
а прибор- проще поискать, где в прокат дают. покупать весьма небюджетно.
 
Sead1995 Дата: Суббота, 22.Окт.2016, 18:54 | Сообщение # 7

Стажер

инженер компании Zdravo

На заявках

Украина

Киев, Украина

Цитата madmac ()

есть дилеры в Киеве и Запорожье, которые, по идее, должны поддерживать гарантию. а как по факту…
уже были стабилизаторы, которые только пишут, что на выходе 220, а так что на входе, то и на выходе.
кроме того, аппарат при съемке просаживает сеть, если длина кабеля на него большая и сечение не то.
опять же в жилых домах утро и вечер- пик потребления жильцами.
а прибор- проще поискать, где в прокат дают. покупать весьма небюджетно.

Я по сути и работаю в компании, которая продает и обслуживает данную продукцию. Но так как полноценного обучения я еще неуспел пройти, то учусь по ходу дела)
Да да, и в этой клинике именно такой стабилизатор и стоит: что на входе — то же и на выходе. И вот такая беда с сетью, даже непонятно что делать.
Ааа, понимаю, слишком дорогие. Будем искать прокат.
Спасибо вам за советы.

 
madmac Дата: Суббота, 22.Окт.2016, 20:12 | Сообщение # 8

Передовик

стоматология и связанное с ней

На заявках

Российская Федерация

СПб

Цитата Sead1995 ()

по сути и работаю в компании, которая продает и обслуживает данную продукцию.

тогда не худо бы купить. пригодится и для рентгенов, и компрессора с автоклавами любят хорошее питание. а контроль перед монтажом нужен.
в старых городах, в центрах, очень часто такая ситуация- старые сети, старые подстанции. да ещё в жилых домах.
так-то надо брать за уши клиента и его электрика, это их задача -обеспечить нормальное ликстричество. но им надо ткнуть в нос параметры сети.

 
МастерБаку Дата: Воскресенье, 23.Окт.2016, 09:25 | Сообщение # 9

инженер

На заявках

Российская Федерация

Нижний Новгород

Цитата madmac ()

покупать весьма небюджетно.

nea
http://ekits.ru/index.php?productID=828
а то начинается — килянусь честный слово ничего не было .

вопрос в том — почему эта штука проводит экспозицию если напряжение мало ?

Сообщение отредактировал МастерБакуВоскресенье, 23.Окт.2016, 09:31

 
madmac Дата: Воскресенье, 23.Окт.2016, 12:56 | Сообщение # 10

Передовик

стоматология и связанное с ней

На заявках

Российская Федерация

СПб

Цитата МастерБаку ()

http://ekits.ru/index.php?productID=828

не то малость. отображает максимум- минимум и все.
фиксация времени, периодичности изменений отсутствует, опять же чувствительность 6 вольт. если сидеть и втыкать в него сутки- пойдет. тогда и мультиметр вполне подходит. сиди с бумажкой в руках, см. на прибор и записывай.
вопрос в периодичности, времени, длительности провалов питания.

Цитата МастерБаку ()

вопрос в том — почему эта штука проводит экспозицию если напряжение мало ?

а пока в пределах+/- 10 вольт он и снимает.
Sead1995, ещё вопрос: время предпрогрева, установленное на плате, соответствует указанному на шильдике моноблока? и вообще замеры делали?

 
Sead1995 Дата: Воскресенье, 23.Окт.2016, 17:47 | Сообщение # 11

Стажер

инженер компании Zdravo

На заявках

Украина

Киев, Украина

Цитата madmac ()

Sead1995, ещё вопрос: время предпрогрева, установленное на плате, соответствует указанному на шильдике моноблока? и вообще замеры делали

Замеры времени прогрева не делал. На заводе они сами устанавливают время прогрева для рентген трубки, которая в комплекте идет и остается только собрать. И сейчас выставленное время прогрева совпадает с тем, которое на моноблоке указано.
А как данные замеры производить?

 
grom1959 Дата: Четверг, 22.Мар.2018, 21:08 | Сообщение # 12

Заглянувший

сервисный инженер

На заявках

Российская Федерация

Спб

С Фона 70 такое-же было,вылечилось покупкой 3 кв стабилизатора
 
Terracorp Дата: Среда, 29.Авг.2018, 08:46 | Сообщение # 13

Участник

Медтехник

На заявках

Российская Федерация

Уфа

Если я правильно понял. спалить аппарат переключением этих переключателей нельзя?
У меня на переключателях позиция, согласно фотографии, на 120 или 240 вольт. Если напряжение ниже 220 на 1 вольт, то загорается ошибка А02. Стабилизатор здесь не причем, так как переключает повышающую обмотку при 215 вольт, а при 216-219 вольт передает напрямую сетевое напряжение на потребители.

То есть у меня там никак не выйдет переключить на 120 вольт, вместо 220? Если исполнение на наши 220, то 110 вольт к нему не относится? Пардон, просто перестраховываюсь, мало ли, аппарат дорогой, отвечать деньгами нет желания.

Сообщение отредактировал TerracorpСреда, 29.Авг.2018, 08:48

 
Terracorp Дата: Вторник, 18.Сен.2018, 19:38 | Сообщение # 14

Участник

Медтехник

На заявках

Российская Федерация

Уфа

Вопрос решен, рентгеновский аппарат Fona X70 выдавал ошибка А02 при напряжении менее 220 вольт. Стабилизатором не лечилось, так как он в диапазоне от 216 до 225 работал в режиме байпас, и по стечению обстоятельств среднее напряжение в сети держалось на 216-219 вольт. Врач запускал рентген, включив чайник в подсобке, тогда срабатывала повышающая обмотка стабилизатора, что тоже не всегда помогало.

Итак, воспользовавшись фото схемы из сообщения №5, переключил программными переключателями (слева второй и третий) в режим работы 220 вольт (хотя до этого стояло на 230 — не работало, видно реально триггер срабатывал при 220, а ниже не хотел работать).

Врач на седьмом небе от счастья. Потому что сервисмены, которые продали и установили ему аппарат, тоже заметили глюки при включении, но предложили купить мощный стабилизатор, чтобы решить проблему. Как я уже писал, стабилизатор не помог, и врач отчаялся, навострив медсестру на манипуляции с чайником, светом и прочим, что выглядело весьма оригинально.

После этого не было ни одной ошибки А02. Если кто будет переключать, на всякий случай не забудьте делать это при обесточенном аппарате, так как я не проверял что будет, если переключать на включенном аппарате.

 

Specifications:

1804/1804767-intraos_70.pdf file (20 Jan 2023)

Accompanying Data:

Blue X IntraOs 70 Medical Equipment PDF Operator’s Manual (Updated: Friday 20th of January 2023 01:20:24 PM)

Rating: 4.1 (rated by 59 users)

Compatible devices: Oxford In-Situ, 930 160, TruTrak+ 3900, ArjoHuntleigh Akron BoBath 5 4632, Pneupac babyPAC 100, AUDIOSTAR PRO, Halo One Series, C-Series.

Recommended Documentation:

Blue X IntraOs 70: Text Version of Document

(Ocr-Read Summary of Contents, UPD: 20 January 2023)

  • 19, IntraOs 70 – Operator’s Manual    Appendix C Exposure Table IntraOs 70 — 70 kVp, 7 mA — Exposure Times in s D Film E Film Digital Small Patient D Film E Film Digital Focus-Film Distance 23 cm D Film E Film Digital Large Pa…

  • 14, IntraOs 70 – Operator’s Manual   1 2 3 4 5 6 7 LINE — SWITCH ON/OFF EXPOSURE HAND-SWITCH CONNECTION EXPOSURE PUSHBUTTON YELLOW LIGHT EXPOSURE HAND-SWITCH CABLE 3.4 Operation Turn on the line voltage supply switch on (below the timer) to bring the system ready Position the image rec…

  • 22, IntraOs 70 – Operator’s Manual   Appendix F Cooling Curves COOLING CURVE OF X-RAY TUBE kJ T (min) COOLING CURVE OF TUBE-HEAD kJ T (min)

  • 8, IntraOs 70 – Operator’s Manual   • Folding arm; useful reach of  55” (140 cm) with 17”¾ (45 cm) extension arm  65” (165 cm) with 27”½ (70 cm) extension arm  72”¾ (185 cm) with 35”½ (90 cm) extension arm  80” (205 cm) with 43”¼ (110 cm) extension arm • The mobile version is composed of:  Mobile base: 29” (74 cm) wi…

  • 10, IntraOs 70 – Operator’s Manual   3. OPERATING INSTRUCTIONS 3.1 Demonstration In order to use of the system for demonstration purposes radiation emission has to be inhibited by disconnecting the supply cables to the tube-head into the wall support or into the timer. Cables to be disconnected are those leaving the connection block towards the tube-h…

  • 9, IntraOs 70 – Operator’s Manual    2.5 AutoSet Timer The timer is working at line voltage of 115 V ± 15%, 50/60 Hz. Mains features are: • Microprocessor controlled functionality • Zero-crossing power switching. • Film speed setting. • Automatic setting of exposure time from 60 ms to 3.2 s through object selection. • Functionality…

  • 18, IntraOs 70 – Operator’s Manual   Appendix B Icons IEC Type B Equipment Compliance to European Community Requirements X-ray On Compliance to Canadian and US Standards Examine Annexed Documentation Line voltage supply On — System Ready Increase Exposure Time (one step)…

  • 24, IntraOs 70 Dental X-ray Equipment Operator’s Manual English Edition USA Version 3.2  

  • 5, IntraOs 70 – Operator’s Manual    1.5 Obligation of the Installer Obligation of the Installer is: • To make sure that the line voltage specified be the Manufacturer of the equipment is available and within the specified range. • For safety reasons verify that a proper switch is available to disconnect from line voltage supply when needed during installa…

  • 7, IntraOs 70 – Operator’s Manual    2.2 Beam Limiting Device • Circular Output Section  Focus Skin Distance: 8.27”(21 cm)  Output Radiation Field: 2.35” diameter (6 cm) • Rectangular Output Section  Focus Skin Distance: 8.27”(21 cm)  Output Radiation Field: 1.8”x 1.81” (3.5 x 4.6 cm) 2.3 Mechan…

  • 13, IntraOs 70 – Operator’s Manual    During the exposure the yellow lamp of X-ray On , on the timer and on the hand-switch, lights and the internal buzzer sounds to indicate radiation emission. As additional safety feature the timer is provided of an independent back-up device (back-up timer and back-up relay) which is going to cut-off radiation i…

Blue X IntraOs 70: Recommended Instructions

DHG397, Bravo Pro, KX-F3100, 2008 Optra

  • HEWI | 1 Rückenstütze Serie 802 LifeSystemBackrest rail Range 802 LifeSystemMontage- und GebrauchsanleitungInstallation and Operating InstructionsNotice de montage et mode d‘emploiInstallatie-en gebruikshandleidingIstruzioni di montaggio e per l’usoInstrucciones de instalación y funcionamientoInstrukcja montażu i obsługi …

    LifeSystem 802 Series 20

  • 1 PEN USER MANUAL Forsteo® FORSTEO 20 micrograms (µg)/ 80 microlitres solution for injection, in pre-filled pen Instructions for use Before you use your new pen, please read the section Instructions for use completely. Follow the directions carefully when using the pen. Also read the package leaflet provided. Do not share your pen or your needles as this may risk transmissi …

    Forsteo 20 5

  • Document Information for:TypeNameRevisionStateDescriptionOriginatorFile ListApproval InformationPerson Action Date and TimeThis page is generated automatically by the GEMS ePDM SystemPrinted documents are for Reference Only and may be out-of-date.Check the database to ensure you have the correct revision.State: RELEASE — Document is released and under formal Change Control. Changes are …

    LOGIQ A3 252

Additional Information:

Operating Impressions, Questions and Answers:

Дентальный рентген Intra OS 70

Поставщики и цены:

Дентальный рентген IntraOS 70 является стоматологическим РТГ аппаратом нового поколения. У рентгена совмещены преимущества высокочасточного и импульсного рентгеновского оборудования:

— фильтрование мягкого излучения — «GRID» улучшенное качество съемки — классической и цифровой

— Высокая надежность — низкие расходы в эксплуатации и сервиса

— Увеличенный срок эксплуатации РТГ головки — пониженная прогреваемость лампы

— Стабилизатор электрического напряжения – регулировка времени проявки снимков

— Простое обслуживание и лёгкая манипуляция

— фотосъёмки на классические РТГ плёнки или через цифровые сенсоры.

Модели прибора:

Передвижной — максимальная длина рукава 1,1 м

Настенный — максимальная длина 1,9 м.

Дентальный рентген IntraOS 70 Технические характеристики

— Потребляемая мощность 70 кВт;

— Сила тока 7 мА;

— Мощность 1,4 кВт;

— Максимальный рабочий цикл 1/60;

— Фокусное расстояние 20 см.;

— Облучаемое поле на конце коллиматора (диаметр) 6 см.

Назначение оборудования — Дентальный рентген Intra OS 70

Похожее оборудование

Отзывы и комментарии:

Blue X IntraOs 70 Operator's Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

IntraOs 70

X-ray Equipment

Operator’s Manual

Blue X Imaging Srl Via Idiomi 1/8-33 20090 Assago ITALY e-mail bluex@bluex.it

loading

Related Manuals for Blue X IntraOs 70

  • Medical Equipment Blue X IntraOs 70 Service & Installation Manual

Summary of Contents for Blue X IntraOs 70

  • Page 1
    IntraOs 70 X-ray Equipment Operator’s Manual Blue X Imaging Srl Via Idiomi 1/8-33 20090 Assago ITALY e-mail bluex@bluex.it…
  • Page 2
    IntraOs 70 X-ray Equipment Operator’s Manual – English Edition USA Version 3.2 December2006 Printed 04/12/2006 11:20 AM Code 69 500 01110 Blue X Imaging is committed to Total Quality. Thank you for notifying us any error found in this document.
  • Page 3: Table Of Contents

    IntraOs 70 – Operator’s Manual Table of Contents Introduction ……………….4 Congratulations…………..4 Purpose ………………4 Equipment Classification…………4 Applicable Standards………….4 Obligation of the Installer…………5 Obligation of the User …………5 Warning ……………..5 Technical Data…………….6 Tube-Head G ……………..6 Beam Limiting Device …………7 Mechanical System…………..7 Weights………………8 AutoSet Timer ……………9…

  • Page 4: Introduction

    The unit is manufactured under a Quality Control System which grants full compliance to specifications. 1.2 Purpose The IntraOs 70 X-ray Equipment is design to fulfil the needs for intra- oral radiography in the general dental practice. The system can be configured for wall, unit or mobile solutions and different types of timers and tube-heads.

  • Page 5: Obligation Of The Installer

    IntraOs 70 – Operator’s Manual 1.5 Obligation of the Installer Obligation of the Installer is: • To make sure that the line voltage specified be the Manufacturer of the equipment is available and within the specified range. • For safety reasons verify that a proper switch is…

  • Page 6: Technical Data

    IntraOs 70 – Operator’s Manual Make sure to carefully manoeuvre the suspension arms to position the tube-head in order prevent harm to the fingers in areas where they may be pinched. Even if compliant to specifications of electromagnetic compatibility, it…

  • Page 7: Beam Limiting Device

    IntraOs 70 – Operator’s Manual 2.2 Beam Limiting Device • Circular Output Section • Rectangular Output Section Focus Skin Distance: Focus Skin Distance: 8.27”(21 cm) 8.27”(21 cm) Output Radiation Field: Output Radiation Field: 2.35” diameter (6 cm) 1.8”x 1.81” (3.5 x 4.6 cm) 2.3 Mechanical System…

  • Page 8: Weights

    IntraOs 70 – Operator’s Manual • Folding arm; useful reach of 55” (140 cm) with 17”¾ (45 cm) extension arm 65” (165 cm) with 27”½ (70 cm) extension arm 72”¾ (185 cm) with 35”½ (90 cm) extension arm 80” (205 cm) with 43”¼…

  • Page 9: Autoset Timer

    IntraOs 70 – Operator’s Manual 2.5 AutoSet Timer The timer is working at line voltage of 115 V ± 15%, 50/60 Hz. Mains features are: • Microprocessor controlled functionality • Zero-crossing power switching. • Film speed setting. • Automatic setting of exposure time from 60 ms to 3.2 s through object selection.

  • Page 10: Operating Instructions

    IntraOs 70 – Operator’s Manual 3. OPERATING INSTRUCTIONS 3.1 Demonstration In order to use of the system for demonstration purposes radiation emission has to be inhibited by disconnecting the supply cables to the tube-head into the wall support or into the timer.

  • Page 11
    IntraOs 70 – Operator’s Manual Index 0.32 0.40 0.50 0.64 0.80 1.00 1.25 1.60 2.00 Speed E type D type D type: Kodak Ultraspeed, Agfa Dentus M2 — E type: Kodak Ektaspeed Plus The timer offers the possibility to correct the exposure time to compensate radiation dose changes due to fluctuations of the line voltage supply, thus granting consistent film blackening.
  • Page 12
    IntraOs 70 – Operator’s Manual To operate the systems switch the timer “on” with the button below the unit. During start-up all the lamps in the panel and the segments of the display, but the yellow light of X-ray On , are blinking for 3 s, to let the operator check that lights in the panel are properly working.
  • Page 13
    IntraOs 70 – Operator’s Manual During the exposure the yellow lamp of X-ray On , on the timer and on the hand-switch, lights and the internal buzzer sounds to indicate radiation emission. As additional safety feature the timer is provided of an independent back-up device (back-up timer and back-up relay) which is going to cut-off radiation in case of failure of the main timer.
  • Page 14: Operation

    IntraOs 70 – Operator’s Manual 3.4 Operation Turn on the line voltage supply switch on (below the timer) to bring the system ready EXPOSURE HAND-SWITCH CONNECTION Position the image receptor LINE-SWITCH where needed and orientate the ON/OFF BLD accordingly. Select the desired exposure time (using the keys or the knob depending on the type of timer connected).

  • Page 15: Safety

    IntraOs 70 – Operator’s Manual 3.5 Safety • Electrical. Trained and qualified technicians only are authorized to remove covers and have access to power circuits. Power supply lines must comply with safety legislation and have ground terminals for protective earth connection.

  • Page 16: Disinfecting

    Refer to Service & Installation Manual for instructions. 4. DISPOSING OF OBSOLETE EQUIPMENT The IntraOs 70 system is made of different materials which include may kinds of metals, iron, aluminium, lead, copper and others, plastic materials, electronic components and dielectric oil in the tube-head.

  • Page 17: Appendix A System Components

    IntraOs 70 – Operator’s Manual Appendix A System Components Article Type Code System Component Catalogue Number Wall Support 93 100 11000 Extension Arm 45 cm 93 100 17100 Extension Arm 70 cm 93 100 17200 Extension Arm 90 cm 93 100 17300…

  • Page 18: Appendix B Icons

    IntraOs 70 – Operator’s Manual Appendix B Icons Compliance to IEC Type B European Community Equipment Requirements Compliance to Canadian and US X-ray On Standards Examine Annexed Line voltage supply Documentation On — System Ready Increase Exposure Off (Disconnected Time (one step)

  • Page 19: Appendix C Exposure Table

    IntraOs 70 – Operator’s Manual Appendix C Exposure Table IntraOs 70 — 70 kVp, 7 mA — Exposure Times in s Small Patient Large Patient 3,20 2,50 2,00 2,60 1,25 1,00 0,80 0,64 0,50 0,40 0,32 2,50 2,00 2,60 1,25 1,00 0,80 0,64 0,50 0,40 0,32 0,25…

  • Page 20: Appendix D Alarm Conditions

    IntraOs 70 – Operator’s Manual Appendix D Alarm Conditions AutoSet Timer Alarm Conditions Code Fault /Error Signal Action Reset A 01 X-ray requested Green lamp System By acknowledgement during cool-down (System Ready) inhibited on the panel or when period flashing…

  • Page 21: Appendix E Identification Labels

    IntraOs 70 – Operator’s Manual Appendix E Identification Labels…

  • Page 22: Appendix F Cooling Curves

    IntraOs 70 – Operator’s Manual Appendix F Cooling Curves COOLING CURVE OF X-RAY TUBE T (min) COOLING CURVE OF TUBE-HEAD T (min)

  • Page 23
    Page intentionally left blank…
  • Page 24
    IntraOs 70 Dental X-ray Equipment Operator’s Manual English Edition USA Version 3.2…

Понравилась статья? Поделить с друзьями:

Интересное по теме:

  • Intex spa e90 ошибка
  • Intex purespa ошибка e98
  • Intex purespa e90 ошибка
  • Intex e81 ошибка
  • Interval expansion minecraft ошибка

  • Добавить комментарий

    ;-) :| :x :twisted: :smile: :shock: :sad: :roll: :razz: :oops: :o :mrgreen: :lol: :idea: :grin: :evil: :cry: :cool: :arrow: :???: :?: :!: